Several Common Troubleshooting Methods for Ultra-low Temperature Freezer in Laboratories
Although the refrigeration system and control circuit of ultra low temperature freezer in laboratory are complex, the common faults are mainly circulating system and circuit part.
Following are several common faults of ultra-low temperature refrigerators and their troubleshooting methods:
1. After power supply is switched on, the air switch of the main power supply is automatically disconnected.
Elimination method: Check whether the compressor of refrigeration system is damaged or not, and replace the parts if damaged. After replacing the fittings, turn on the machine to check whether it is normal. The fault is usually caused by unstable power supply, because the * stage starts first, so the damage is usually in stage. In this case, as long as the delay regulated power supply is installed, the fault will not occur again.
2. Stop working before the temperature in the box reaches the set temperature.
Exclusion: Although the temperature in the box has not reached the set temperature, the panel display temperature has reached the set temperature. This is due to the damage of the thermistor which detects the temperature and the need to replace the parts. If there is no original accessories, a set of temperature control system can be re-installed to eliminate the failure.
3. Poor refrigeration effect may be refrigerant leakage
4. If the ultra-low temperature refrigerator does not start in winter, it may be that the thermostat regulation value is too small.
We are a ultra low temperature freezer manufacturer. Welcome to choose our products.
Похожие рекомендации
-
Structural Characteristics of Industrial Recirculating Chiller
1406In the chemical reaction, there are strict requirements for the reaction temperature. It is necessary to raise the temperature to an appropriate temperature to carry out the chemical reaction smoothly. After the reaction, the temperature in the re...
Посмотреть детали -
The Application of Industrial Chillers
1443The chiller machine is needed in a lot of industries, basically chemical, medical, pharmaceutical industries are involved, then what industry can Wuxi Lneya chiller meet in the end? It is known that we are one of the top chiller manufacturers. Th...
Посмотреть детали -
Introduction to the structure and application of industrial refrigerators
1555Industrial refrigerators are usually called industrial refrigerator boxs, and low-temperature freezers.Temperature range: -150℃~-20℃. Industrial refrigerators can be divided into vertical and horizontal types. Most of them use imported compressors...
Посмотреть детали -
Применение системы контроля температуры газового охлаждения в оптическом модуле
1571The gas cooling temperature control system has a wide range of temperatures and can be widely used to meet the requirements of more production and engineering environments. It is applicable to reliability tests such as chip, microelectronic devi...
Посмотреть детали
LNEYA Промышленные чиллеры Производитель Поставщик