LNEYA Low temperature thermostatic bath instructions
Low temperature thermostat is widely used in petroleum, chemical industry, electronic instruments, physical, chemical and biological engineering, medicine and health, life sciences, food industry, physical testing and chemical analysis and research departments, universities, enterprises and production quality inspection departments, for the user to work to provide a heat source temperature field controlled, uniform and constant constant temperature test, or test on the test samples or products, but also can be used as direct heating or cooling and heating or cooling cooling or heating.
Low temperature thermostatic bath is widely used to provide a user with a controlled and uniform temperature field source, a constant temperature test for a test sample or a produced product be used as a heat source or cold source for direct heating or cooling or auxiliary heating or cooling.
1. The liquid medium must be added first in the low temperature or constant temperature bath before using, otherwise, the heater will be damaged when the electricity is working. The liquid surface of the medium should be lower 30mm than the working plate.
2. The selection of liquid medium in low temperature and constant temperature bath should be in accordance with the following principles:
3. When the working temperature is at 5℃~85 ℃, choose water as the liquid medium.
4. When the working temperature is at 85℃~95 ℃, choose 15% glycerol solution as the liquid medium.
5. When the working temperature is below 5 centigrade, choose alcohol or frozen liquid as the liquid medium.
6. When the cryogenic thermostat switch is closed, the refrigeration switch should be opened for a 10~20 minutes interval (depending on the ambient temperature).
7. Low temperature constant temperature bath power: 220V50Hz, the power of the power supply should be greater than the total power of the instrument. The power supply must have a good “grounding” device.
8. The low temperature and constant temperature bath should be placed in the dry and ventilated place, and there are no obstacles in the 300mm around the instrument. When low temperature and constant temperature bath temperature is low, don’t put hands into the slot, in order to prevent frost injury. After using, all switches should be closed and the cut the power off. Clean the Low Temperature Thermostat Bath frequently and keep the work table and the operation panel clean.
Похожие рекомендации
-
Water cooled low-temperature circulators in the testing industry of new energy vehicles
1586The development of the new energy vehicle industry is relatively rapid. Driven by technological innovation, it is believed that the new energy vehicle industry is facing opportunities. However, with the development of the new energy vehicle indu...
Посмотреть детали -
How To Vacuum The Ultra-Low Temperature Freezer Refrigeration System?
1682LNEYA ultra-low temperature freezer indoor and outdoor connection has been completed, observe the oil pump oil mark indication to see if there is enough oil, and then start the vacuum pump to see if it is normal. Then unscrew the large valve back ...
Посмотреть детали -
New energy vehicle battery motor test motor oil chiller and liquid chiller
1767New energy vehicle battery testing chiller
Посмотреть детали -
Точки внимания для ежедневной проверки оборудования для контроля низких температур при минус 30℃
1573Оборудование для поддержания низких температур при минус 30 градусах Цельсия - это холодильный агрегат или еще его называют хладоагрегат. Отличие от чиллера заключается в температурном диапазоне. Это широко используемое низкотемпературное оборудование в промышленно...
Посмотреть детали
LNEYA Промышленные чиллеры Производитель Поставщик