Как устранить сигнал тревоги высокого давления в промышленном ультранизкотемпературном морозильнике?
During use, if a fault is not solved in time, it will affect the use of industrial ultra-low temperature freezer. How to solve the problem of high-pressure alarm?
If the water-cooled industrial ultra-low temperature freezer presents a high-pressure alarm, it is mainly related to the cooling water system. The water-cooled industrial ultra-low temperature freezer provides circulating cooling water through a cooling tower. If you forget to open the cooling water valve during use, If the cooling water cannot be circulated to the chiller, it will show high pressure problems. In addition, check whether the cooling water tower is defective and cannot safely and stably deliver the cooling water.
The cooling water flow is too small, or the temperature is too high and the heat dissipation is not good. This problem will also occur. If the cooling water flow is small, it is necessary to check whether the pipe size of the supporting device matches the size of the industrial ultra-low temperature freezer group, or whether the water valve can be opened to a large position; if it is an air-cooled industrial ultra-low temperature freezer, we need to look at the ambient temperature of the device If it can be exceeded, the heat must be exhausted in time to ensure that the industrial ultra-low temperature freezer unit is in an air circulating environment.
Industrial ultra-low temperature freezers are used for too long, and it is easy to accumulate dirt and other debris on the pipe wall or condenser copper pipes, which will inevitably affect the refrigeration cycle system. Therefore, we regularly provide industrial ultra-low temperature freezers according to the maintenance time recommended by the manufacturer Do maintenance cleaning work (cleaning the cooling water system, cleaning the condenser, changing the filter, etc.).
Check whether the refrigerant is excessive. The industrial ultra-low temperature freezer unit will be filled with the appropriate amount of refrigerant before leaving the factory. Therefore, if there is no leakage of the refrigerant, the user does not need to charge the refrigerant. Too much refrigerant will only affect the unit’s failure. use.
Whether the opening of the shrink valve is too small, for this, everyone can gradually adjust the opening of the shrink valve to eliminate the problem.
Похожие рекомендации
-
The Application of Industrial Chillers
1440The chiller machine is needed in a lot of industries, basically chemical, medical, pharmaceutical industries are involved, then what industry can Wuxi Lneya chiller meet in the end? It is known that we are one of the top chiller manufacturers. Th...
Посмотреть детали -
Описание метода обслуживания машины с холодной и горячей температурой одной пресс-формы
1842В процессе использования горячих и холодных интегрированных температурных машин для пресс-форм, необходимо регулярно проводить техническое обслуживание и ремонт. Затем, LNEYA поделится с вами методом обслуживания горячей и холодной интегрированной машины температуры формы...
Посмотреть детали -
Attention points for the use of laboratory chiller manufacturers’ equipment
1429In the laboratory, a wide range ofexperimental instruments are widely used. When the experimental instruments arein operation, they are inseparable from the high-low temperature circulatingdevice refrigeration heating equipment. What should be ...
Посмотреть детали -
Maintenance Method Of Ultra Low Temp Chiller Refrigeration System
1692The ultra low temp chiller refers to the refrigeration unit whose chilled water temperature reaches below minus 35℃. The unit adopts two-stage compression refrigeration, and the combination of the scroll compressor and the piston compressor is use...
Посмотреть детали
LNEYA Промышленные чиллеры Производитель Поставщик