Как устранить сигнал тревоги высокого давления в промышленном ультранизкотемпературном морозильнике?
During use, if a fault is not solved in time, it will affect the use of industrial ultra-low temperature freezer. How to solve the problem of high-pressure alarm?
If the water-cooled industrial ultra-low temperature freezer presents a high-pressure alarm, it is mainly related to the cooling water system. The water-cooled industrial ultra-low temperature freezer provides circulating cooling water through a cooling tower. If you forget to open the cooling water valve during use, If the cooling water cannot be circulated to the chiller, it will show high pressure problems. In addition, check whether the cooling water tower is defective and cannot safely and stably deliver the cooling water.

The cooling water flow is too small, or the temperature is too high and the heat dissipation is not good. This problem will also occur. If the cooling water flow is small, it is necessary to check whether the pipe size of the supporting device matches the size of the industrial ultra-low temperature freezer group, or whether the water valve can be opened to a large position; if it is an air-cooled industrial ultra-low temperature freezer, we need to look at the ambient temperature of the device If it can be exceeded, the heat must be exhausted in time to ensure that the industrial ultra-low temperature freezer unit is in an air circulating environment.
Industrial ultra-low temperature freezers are used for too long, and it is easy to accumulate dirt and other debris on the pipe wall or condenser copper pipes, which will inevitably affect the refrigeration cycle system. Therefore, we regularly provide industrial ultra-low temperature freezers according to the maintenance time recommended by the manufacturer Do maintenance cleaning work (cleaning the cooling water system, cleaning the condenser, changing the filter, etc.).
Check whether the refrigerant is excessive. The industrial ultra-low temperature freezer unit will be filled with the appropriate amount of refrigerant before leaving the factory. Therefore, if there is no leakage of the refrigerant, the user does not need to charge the refrigerant. Too much refrigerant will only affect the unit’s failure. use.
Whether the opening of the shrink valve is too small, for this, everyone can gradually adjust the opening of the shrink valve to eliminate the problem.
Похожие рекомендации
-
Охлаждающая система отопления дистилляционной колонны
1097Дистилляционные колонны и процессы в дистилляционных колоннах действительно требуют использования охладителей и систем отопления для поддержания температурного режима в ходе процесса, обеспечивающего эффективное разделение и очистку. Верхняя температура ...
Посмотреть детали -
Importance Of The Low Temperature Chamber
897The low temperature chamber is a large cryogenic or high-cryogenic test device, which usually separates the refrigerator from the test part and makes the test part into a room type, so the staff can work in it. Such a test room is called a low tem...
Посмотреть детали -
Refrigeration heating circulator list resonance fault description
1228Refrigeration heating circulator leaderboard works: When the temperature of the reactor, reactor, rotary evaporator, extraction equipment, etc. of the refrigeration heating circulator needs to be rapidly increased or lowered, the equipment can qui...
Посмотреть детали -
Почему для шпинделей станков требуются охладители?
903The main reasons why machine tool spindles require a chiller are as follows: Temperature Control: During the machining process, the spindle generates a large amount of heat due to high-speed rotation. If not controlled, this heat ...
Посмотреть детали