KRYZ Frequency conversion series
Control Temperature and Flow -40℃~100℃- Мощность охлаждения0.5kW~10kW
- Мощность нагрева0.5kW~10kW
- ХладагентR134A/R1234yf
- Вес115KG~300KG
Air-cooling & Water-cooling
Детали конфигурации
Модель | KRYZ-475 KRYZ-475W |
KRYZ-4A15W | KRYZ-4A25W | |
Диапазон температур | -40℃~+100℃ | |||
Точность контроля температуры | ±0.1℃ | ±0.1℃ | ±0.1℃ | |
Обратная связь по температуре | Pt100 | Pt100 | Pt100 | |
Отображение температуры | 0.01k | 0.01k | 0.01k | |
Flow output | 1~25 L/min | 1~25 L/min | 1~25 L/min | |
About traffic description | When the temperature is below -30 degrees, the flow rate is controlled within 80% of the maximum value. | |||
Flow control accuracy | ±0.1 L/min | ±0.1 L/min | ±0.1 L/min | |
Pressure display | Using pressure sensor, the pressure is displayed on the touch screen, and the pressure control can be adjusted from 0.2bar to 2.5bar | |||
Heating capacity kW | 7.5 | 15 | 25 | |
Cooling capacity kW | 100℃ | 7.5 | 15 | 25 |
20℃ | 7.5 | 15 | 25 | |
0℃ | 7.5 | 15 | 25 | |
-20℃ | 4.5 | 10 | 16 | |
-35℃ | 1.8 | 4 | 6.5 | |
Компрессор | Inverter scroll compressor | |||
Метод дросселирования | Электронный расширительный клапан | |||
Маслоотделитель | Эмерсон | |||
Фильтр-осушитель | Danfoss | |||
Испаритель | Пластинчатый теплообменник | |||
Циркуляционный насос | Variable frequency magnetic drive pump | |||
Вход, дисплей | 7-inch color touch screen, PLC controller | |||
Programming | 5 programs can be compiled, each program can compile 40 steps | |||
Communications | Modbus RTU protocol RS485 interface, optional CAN communication bus, Ethernet TCP/IP protocol | |||
Защита безопасности | It has self-diagnosis function; refrigerator overload protection; high pressure switch, overload relay, thermal protection device, low liquid level protection, high temperature protection, sensor fault protection and other safety protection functions. | |||
System | The entire system is a fully closed system. There will be no water vapor at high temperatures and no absorption of moisture in the air at low temperatures. The system will not cause pressure to rise due to high temperatures during operation, and will automatically replenish heat transfer media at low temperatures. | |||
Хладагент | R404A/ R507C | |||
Размер интерфейса | G3/4 | G3/4 | G3/4 | |
Water cooled @25℃ | 2500L/H ZG3/4 |
4000 Л/ЧАС ZG1 |
6000 Л/ЧАС DN32 |
|
Конденсатор с водяным охлаждением | The water equipment condenser of this product adopts plate heat exchanger, which requires water supply from chiller or low hardness tap water, and requires water purification filter | |||
Air cooled condenser | Microchannel Heat Exchanger | / | / | |
Power supply 380V50HZ | 14 кВт | 25 кВт | 38 кВт | |
Размеры см | 55*85*175 | 70*100*175 | 80*120*185 | |
Вес | 280 кг | 360 кг | 620kg | |
Standard | Automatic antifreeze filling system, automatic liquid recovery system | |||
Дополнительно | 220V 60HZ three-phase 400V 50HZ three-phase 440V 60HZ three-phase | |||
Дополнительно | Temperature extended to -40℃~+135℃ | |||
Ремарка | The external circulation pipeline should have a diameter of DN15 or DN20. The pipeline needs to be well insulated, and high-temperature resistant rubber and plastic insulation pipes can be used. | |||
Материал оболочки | Cold rolled sheet spraying (standard color 7035) |
Модель | KRYZ-475W/2T | KRYZ-4A15W/2T | KRYZ-4A25W/2T | |
Диапазон температур | -40℃~+100℃ | |||
Точность контроля температуры | ±0.1℃ | ±0.1℃ | ±0.1℃ | |
Обратная связь по температуре | Pt100 | Pt100 | Pt100 | |
Отображение температуры | 0.01k | 0.01k | 0.01k | |
Flow output | 1~25 L/min | 1~25 L/min | 1~25 L/min | |
About traffic description | When the temperature is below -30 degrees, the flow rate is controlled within 80% of the maximum value. | |||
Flow control accuracy | ±0.1 L/min | ±0.1 L/min | ±0.1 L/min | |
Fluid output groups | 2 groups | 2 groups | 2 groups | |
Pressure control per group | 0.2bar~2.5bar | 0.2bar~2.5bar | 0.2bar~2.5bar | |
Flow Pressure Description | The device can control flow or pressure independently. Different solutions are required for simultaneous control (customization required) | |||
Heating capacity kW | 7.5*2 | 10*2 | 15*2 | |
Cooling capacity kW | 100℃ | 7.5*2 | 15*2 | 25*2 |
20℃ | 7.5*2 | 15*2 | 25*2 | |
0℃ | 7.5*2 | 15*2 | 25*2 | |
-20℃ | 4.5*2 | 8.5*2 | 14*2 | |
-35℃ | 1.8*2 | 4*2 | 6.5*2 | |
Компрессор | Inverter scroll compressor | |||
Метод дросселирования | Электронный расширительный клапан | |||
Маслоотделитель | Эмерсон | |||
Фильтр-осушитель | Danfoss | |||
Испаритель | Пластинчатый теплообменник | |||
Циркуляционный насос | Variable frequency magnetic drive pump | |||
Flow rate, pressure | Electromagnetic flowmeter, pressure value 4~20mA | |||
Вход, дисплей | 7-inch color touch screen, PLC controller | |||
Communications | Modbus RTU protocol RS485 interface, optional CAN communication bus, Ethernet TCP/IP protocol | |||
Защита безопасности | It has self-diagnosis function; refrigerator overload protection; high pressure switch, overload relay, thermal protection device, low liquid level protection, high temperature protection, sensor fault protection and other safety protection functions. | |||
System | The entire system is a fully closed system. There will be no water vapor at high temperatures and no absorption of moisture in the air at low temperatures. The system will not cause pressure to rise due to high temperatures during operation, and will automatically replenish heat transfer media at low temperatures. | |||
Хладагент | R404A/R507C/R125/N40 four refrigerants can be used | |||
External output interface size | There are 2 sets of inlet and outlet ports, G3/4 | |||
Water cooled @25℃ | 4000 Л/ЧАС ZG1 |
8000 Л/ЧАС DN32 |
12000 Л/ЧАС DN40 |
|
Конденсатор с водяным охлаждением | The water equipment condenser of this product adopts plate heat exchanger, which requires water supply from chiller or low hardness tap water, and requires water purification filter | |||
Размеры см | 80*120*185 | 100*150*185 | 200*145*205 | |
Power supply 380V50HZ | 28 кВт | 45 кВт | 70kW | |
Вес | 660KG | 850 кг | 1350kg | |
Standard | Automatic antifreeze filling system, automatic liquid recovery system | |||
Дополнительно | 220V 60HZ three-phase 460V 60HZ three-phase | |||
Дополнительно | Temperature extended to -40℃~+135℃ | |||
Ремарка | The external circulation pipeline should have a diameter of DN15 or DN20. The pipeline needs to be well insulated, and high-temperature resistant rubber and plastic insulation pipes can be used. | |||
Материал оболочки | Cold rolled sheet spraying (standard color 7035) |
Rated test conditions: dry bulb temperature 20℃; wet bulb temperature 16℃. Inlet water temperature 20℃; outlet water temperature 25℃.
The data in this table is for reference only. Please refer to the unit nameplate parameters.
If you need special non-standard customization, please contact us

The equipment can meet the direct cooling/direct heating for battery pack and vehicle testing and testing. This equipment is mainly used for cooling and heating capacity testing. It uses variable frequency compressor with adjustable output; variable frequency fan with adjustable speed and heat dissipation; with high-precision temperature, pressure, flow sensors and built-in software, it can calculate heat exchange capacity in real time, automatically recycle and refill refrigerators, etc. It can meet the needs of battery packs of different sizes.
Вас также могут заинтересовать эти продукты
Инвертор LTZ серии M
APPLICATIONS Suitable for application to the target temperature control object already has a cold storage tank and a circulating pump, and the medium is transported into the LTZ M series refrigeration temperature control unit through th…
LN -60°C~ -10°C
适用范围 Характеристики продукта 产品参数 Параметры продукта 行业应用 ПРИМЕНЕНИЕ 航空航天材料|试验装置控温解决方案 在航空航天领域,材料的性能直接关系到飞行器的安全、可靠性和...
LT -80℃~-20℃
It can guarantee 24-hour continuous operation Circulating refrigeration of ethylene glycol mixed with water mixture 1. Can use ethylene glycol mixed with water for circulating refrigeration, saving resources 2.Cir…