Failures of Heating Cooling Circulators
Сайт heating cooling circulator can be used in many chemical pharmaceuticals and laboratories. Then, how to solve the failures of the heating cooling circulator?
Failure of circulating pump
The failure of the circulating pump is generally the heat overload of the circulating pump. There are 4 main reasons for this failure:
1. The external circulation pipeline is too long. It is suggested to shorten the external circulation pipeline, reduce the bending of the external circulation pipeline, and then click the Reset button of the electric cabinet surface heat relay;
2. The circulating pipe is blocked. It is suggested to check the degree of cleanliness of the circulating pipe, install a filter, and then click the Reset button of the electric cabinet heat relay;
3. The circulation pump voltage is too low. It is suggested to measure whether the circulation pump voltage is in line with the prescribed range, improve the voltage environment, and then click the Reset button of the electric cabinet heat relay;
4. The regulating current of the heat relay is too small. It is suggested to measure the pump current, compare the heat relay range, turn up the heat relay current properly with the cross screwdriver, and then click the Reset button of the electric cabinet heat relay.
Failure of sensor
The great possibility of sensor failure is the wiring error. At this time, check the wiring in time, and we should pay attention to two kinds of situations, one is the sensor is short-circuited or the temperature is < -160 ℃, the other is the sensor is open circuited or the temperature is > 349℃.
If the sensor is short-circuited or the temperature is <-160 ℃, it may be the fault of the sensor itself. At this time, change the sensor (there is a standby sensor in the equipment line slot), if the sensor connection is wrong, it is suggested to check the connection condition (the control circuit diagram) to correct.
If the sensor is open circuited or the temperature is > 349 ℃, it should be the fault of the sensor itself. At this time, replace the sensor in time (there is a standby sensor in the equipment line slot). It is necessary to pay attention to whether the temperature is too high and beyond the range. If the equipment is out of electricity, check the power supply of the equipment.
The above two failures of the heating cooling circulator are relatively simple. I believe that you can quickly solve these failures with these tips.
Похожие рекомендации
-
Осаждение тонких пленок подложки температура контроля оборудования чиллер использования окружающей среды
1638Среда использования для охладителя охлаждения осаждения тонкой пленки вафли
Посмотреть детали -
Турнир по бадминтону LNEYA 2020 успешно завершился
1736In order to enrich employees' spare time, enhance cohesion among employees, create a sunny, energetic and passionate working atmosphere, and demonstrate the company's positive, healthy and upward spirit, LNEYA held the third "Guanya Cup" badminton...
Посмотреть детали -
Several Attention Aspects of Ultra Low Temperature Freezer
1785The performance of ultra low temperature freezer will be reduced naturally due to long-period application in all fields. Thus, Guanya Refrigeration Technology Co., Ltd, leader in ultra low temperature freezer manufacturers has to remind every user...
Посмотреть детали -
Installation points of cryopumps in cryogenic refrigerators
1899In the construction of the cryogenic refrigerator, the cryopump is a must. The low temperature freezer manufacturer LNEYA tells everyone that the cryopump should be operated according to the requirements when it is installed. Installation faciliti...
Посмотреть детали
LNEYA Промышленные чиллеры Производитель Поставщик